Long-read sequencing

Circular consensus sequencing in nanowells: inserts sheared to ~15–20 kb are made into circles, read in many passes by one polymerase, and combined into reads usually above Q30.

kb-scale readsSimulator readyCore simulatorEngine: Sequencing & library prep
Choose a sub-technique

Our ruleSimulations prepare you for the bench but don't replace hands-on practice.

Sub-technique you chose

PacBio HiFi

Circular consensus sequencing in nanowells: inserts sheared to ~15–20 kb are made into circles, read in many passes by one polymerase, and combined into reads usually above Q30.

How it differs from the standard method

  • Insert size is chosen at shearing and size selection; longer means fewer passes.
  • Loading targets about one molecule per well; single-loaded wells top out near 37%.
  • DNA damage repair matters: damage stalls the polymerase and costs passes.
  • Kinetics of incorporation also report base modifications such as 5mC.

1One molecule, read end to end

Short-read platforms read millions of clonal clusters a few hundred bases at a time, then rebuild genomes computationally. Long-read platforms read single molecules thousands to hundreds of thousands of bases long, with no amplification.

HiFi

Circular consensus

A polymerase in a nanowell copies a circular template over and over; the passes are combined into one highly accurate read of 10–25 kb.

Nanopore

Strand through a pore

A motor protein feeds DNA through a protein pore and the ionic current is decoded into bases; reads are as long as the molecules, sometimes over 100 kb.

N50

Read length

Half of all bases sit in reads at least this long; the standard summary of a long-read run.

2What sets length, accuracy and yield

  • DNA integrity first: you can shear long DNA shorter but never make short DNA longer. Gentle extraction and handling decide the ceiling.
  • HiFi trades insert length against passes. A polymerase reads a roughly fixed total length per well, so longer inserts get fewer laps and lower consensus quality.
  • Loading follows Poisson statistics in nanowells. At most about 37% of wells hold exactly one molecule, at an average of one per well. Overloading creates multi-loaded wells that are discarded.
  • Pores capture molecules, not mass, so nanopore libraries are loaded in fmol calculated from the real fragment length. Too much long DNA clogs pores and shortens the run.
  • Nanopore accuracy depends on the basecalling model: faster models are noisier, and the raw signal can be re-basecalled later.
  • Short fragments are captured or loaded preferentially, so size selection protects read length.

3Choosing between them

HiFi reads are both long and accurate (typically Q30 or better), which suits small-variant calling, phasing and assembly. Their length is limited by the trade against passes.

ZMW (zero-mode waveguide)

A nanowell small enough that only the polymerase at its floor is illuminated, so one molecule is watched at a time.

SMRTbell

A circular template made by ligating hairpin adapters to both ends of a double-stranded insert.

Pass / subread

One traversal of the insert by the polymerase; several are combined into the consensus read.

Polymerase read length

The total sequence a polymerase produces before it stops; it bounds the number of passes.

Pore occupancy

The share of active nanopores reading a strand at a given moment rather than waiting for one.

Read N50

The length such that half of all sequenced bases are in reads at least that long.

Spanning read

A read that covers a feature and anchors in unique sequence on both sides; needed to size repeats and resolve structural variants.

Basecalling model

The neural network that turns raw nanopore signal into bases; fast, high-accuracy and super-accuracy variants trade compute for accuracy.

When to use it

  • Sizing repeat expansions and resolving structural variants, segmental duplications and complex rearrangements.
  • De novo genome assembly, phasing of haplotypes, and closing gaps in reference genomes.
  • Full-length transcript isoforms (long-read RNA or cDNA sequencing).
  • Native methylation and other base modifications without bisulfite conversion.
  • Whole-plasmid or whole-amplicon verification in a single read.

Compared with alternatives

Short-read sequencing →

Cheaper per base and very accurate; prefer it for counting applications and small variants in unique sequence.

Sanger sequencing →

One accurate ~800 bp read per reaction; still the quick check for one junction or variant.

Whole-genome sequencing →

The application layer: germline, somatic and assembly analyses can be run on long or short reads.

Library construction →

Short-read library prep; long-read libraries keep fragments long and are usually not amplified.

Sub-techniques

Choosing one here switches the whole page, including the protocol and the lab, to it.

Circular consensus sequencing in nanowells: inserts sheared to ~15–20 kb are made into circles, read in many passes by one polymerase, and combined into reads usually above Q30.

  • Insert size is chosen at shearing and size selection; longer means fewer passes.
  • Loading targets about one molecule per well; single-loaded wells top out near 37%.
  • DNA damage repair matters: damage stalls the polymerase and costs passes.
  • Kinetics of incorporation also report base modifications such as 5mC.