SPR & BLI
Analyte flows through microfluidic channels over a gold sensor chip (e.g. CM5 dextran). Very sensitive and well suited to detailed kinetics, but needs careful referencing and regeneration.
Our ruleSimulations prepare you for the bench but don't replace hands-on practice.
Surface plasmon resonance (Biacore)
Analyte flows through microfluidic channels over a gold sensor chip (e.g. CM5 dextran). Very sensitive and well suited to detailed kinetics, but needs careful referencing and regeneration.
How it differs from the standard method
- Amine coupling at pH below the ligand's pI.
- Keep Rmax low (~50–150 RU) for kinetics.
- Flow 30–100 µL/min to avoid mass transport.
1Watching binding happen in real time
Both methods immobilise one partner (the ligand) on a sensor surface and flow or dip the other (the analyte) over it. As analyte binds, mass builds up at the surface and an optical signal changes.
Surface plasmon resonance
Light reflected from a gold film under the sensor chip: the resonance angle shifts as the refractive index next to the gold changes. 1 RU ≈ 1 pg/mm² of protein.
Bio-layer interferometry
White light reflected from two layers at the tip of a fibre-optic biosensor: binding thickens the layer and shifts the interference pattern.
2From curves to constants
For a simple 1:1 interaction, dR/dt = ka·C·(Rmax − R) − kd·R. Each concentration gives a curve whose association rate (kobs = ka·C + kd) and dissociation rate (kd) are fitted globally across the series.
3Artefacts that look like binding
- Bulk refractive-index jumps from buffer mismatch: square steps at the start and end of each injection. Removed by subtracting a reference surface.
- Mass-transport limitation: with too much ligand or too little flow, analyte binds faster than it is delivered, so curves look linear and sluggish.
- Incomplete regeneration: analyte left from previous cycles raises baselines and lowers capacity.
- Too short a dissociation: if the signal barely drops, kd (and so KD) is undefined.
Key terms
Practise these as flashcards →Ligand is immobilised on the surface; analyte is in solution.
Association rate constant, M⁻¹s⁻¹.
Dissociation rate constant, s⁻¹; complex half-life = ln2 / kd.
Equilibrium dissociation constant = kd/ka, in M. Lower means tighter binding.
Maximum response if every ligand site is occupied; scales with ligand density and analyte/ligand mass ratio.
Subtracting both a reference surface and buffer-only injections to remove bulk effects and drift.
When to use it
- Ranking antibodies, nanobodies or drug candidates by affinity and off-rate.
- Measuring kinetics label-free with small sample amounts.
- Checking that a purified protein is active (binds its partner).
Compared with alternatives
Gives KD, stoichiometry and enthalpy in solution without immobilisation, but no kinetics and much more sample.
KD in solution with tiny volumes; no on/off rates.
Endpoint, labelled, high-throughput; doesn't give kinetics.
Solution binding of a small labelled ligand; good for competition screens.
Sub-techniques
Choosing one here switches the whole page, including the protocol and the lab, to it.
Analyte flows through microfluidic channels over a gold sensor chip (e.g. CM5 dextran). Very sensitive and well suited to detailed kinetics, but needs careful referencing and regeneration.
- Amine coupling at pH below the ligand's pI.
- Keep Rmax low (~50–150 RU) for kinetics.
- Flow 30–100 µL/min to avoid mass transport.